Products  |  Thermodynamics, Heating and Sanitary Systems  |  Internal Combustion Engines  |  CT 110  |  CT 100.11
CT 100.11 Exhaust Gas Calorimeter for CT 110 Show data sheet (pdf file)
Technical Description

Determination of the thermal exhaust gas losses is essential when calculating an energy balance for internal combustion engines. Calorimetric measurement is an established method of doing this. It involves a largely complete and loss-free heat exchange between the exhaust gas and a cooling medium.
The exhaust gas calorimeter consists of an insulated stainless steel tank, through which the exhaust gas flows from bottom to top. While doing this, the exhaust gas gives up its heat almost completely to a finned pipe with cooling water flowing through it. The pipe is arranged in loops to achieve the maximum possible heat exchange area. Relevant temperatures (water and exhaust gas inlet and outlet) and the volumetric flow of the water are recorded electronically and displayed digitally using a measuring amplifier. The measuring amplifier is connected to the CT 110 test stand using a data cable. This allows the measured data from the CT 110 to be stored and processed using PC data acquisition.
The CT 100.11 is connected to a test engine (CT 100.20 - CT 100.23) using a heat-resistant exhaust gas hose.

 

Learning Objectives / Experiments

- Determination of exhaust gas thermal output power

  given up
- Determination of specific heat capacity of exhaust

  gas

Features

* Counterflow heat exchanger for calorimetric analysis

  of exhaust gases from internal combustion engines1
* PC data acquisition for calorimetric test

Specification

[1] Determination of the amount of heat contained in the exhaust gas from test engines
[2] Calorimeter consisting of finned pipe heat exchanger and insulated tank
[3] Instrumentation: 4 temperature sensors, flow meter
[4] Measuring amplifier with digital displays
[5] Connection between engine and calorimeter using exhaust gas hose
[6] LabVIEW software included in CT 110 software
[7] Connection between measuring amplifier and CT 110 using data cable
[8] Cold water connection and drain required

 

Technical Data

Calorimeter:
- insulated, stainless steel

Finned pipe heat exchanger:
- heat exchange area on exhaust gas side: 1.169m²
- heat exchange area on water side: 0.164m²
 
Measuring ranges:
- exhaust gas temperature: 2x 0…600°C
- water temperature: 2x 0…200°C
- flow rate: 0...600L/h

 

Dimensions and Weight
l x w x h : 570 x 570 x 1300 mm (basic unit)
l x w x h : 370 x 300 x 150 mm (measuring amplifier)
Weight : approx. 105 kg
Connections
230V, 50/60Hz, 1 phase or 120V, 50/60Hz, 1 phase or 120V, CSA, 1 phase
Cold water supply 200L/h and drain
Scope of Delivery
1 calorimeter
1 measuring amplifier with digital displays
1 exhaust gas hose
1 data cable for connection to CT 110
1 instruction manual
Order Details

063.10011  CT 100.11  Exhaust Gas Calorimeter

                    for CT 110


© 2005 G.U.N.T. Gerätebau GmbH